ClearanceBrief
Recall Z-0571-2008 ยท NQN Microscope, Automated, Image Analysis, Immunohistochemistry,Operator Intervention,Nuclear Intensity & Percent Positivity

Ventana Medical Systems Inc

Ventana Image Analysis System (VIAS) - PATHWAY anti-HER-2/neu (4B5) primary antibody with software version 2.3.0.2, Model Number: VIAS-799-10000-00, Pathology Image Analysis System, Ventana Medical Systems, Inc., Tucson, AZ 85755

DateMar 18, 2008 (initiated Jun 26, 2007)
StatusTerminated
ReasonIncomplete upgrade: The software upgrade from February 1, 2007 may not have been completed. This software upgrade adds the algorithm (Slidetype) for the PATHWAY anti-HER-2/neu (4B5) antibody to the VIAS system.
Root causeSoftware change control
ActionBeginning in 6/07, Ventana Medical Systems, Inc., made phonecalls to labs asking to confirm software version, date of upgrade if available and confirmation of whether site uses 4B5 and Imaging together. A phone script was provided. Customers with systems not ugraded were to upgrade before further use of system. A notification letter dated 8/17/2007 was later sent to VIAS customers advising that the software upgrade from February 1 , 2007 may not have been completed on their system. This software upgrade adds the algorithm (Slidetype) for the PATWAY anti-HER-2/neu (4B5) antibody to their VIAS
Quantity77 devices
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061613VENTANA IMAGE ANALYSIS SYSTEM - PATHWAY HER2 (4B5)
Tripath Imaging, Jan 10, 2007