Recall Z-0293-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Toshiba American Med Sys Inc

Toshiba-Interventional Angiography System, Infinix i, Model DFP-8000D, equiped with Digital Radiography System with Software version 3.4er000, Toshiba America Medical Systems, Inc., Tustine, CA 92780

DateNov 27, 2007 (initiated Oct 1, 2007)
StatusTerminated
ReasonSystem Lock-up: When performing DAS One Shot, the system may lockup due to a software timing error.
Root causeSoftware design
ActionNotification letters (Urgent: Medical Device Correction-dated October 1, 2007) with an attached Customer Response form were sent to customers beginning on October 2, 2007. Customers were notified this action was being taken as a result of an investigation that resulted from a software issue that was found done by the manufacturer. Customers were informed that a Toshiba Customer Service Engineer would contact them to schedule a software upgrade when the software becomes available. If customers have any questions they were instructed to contact their or call (800) 421-1968 and ask for the Regula
Quantity23
DistributionNationwide-including states of AL, AR, FL, GA, MI, NY, OH, PA, TX & WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052884MODEL DFP-8000D/FPD DIGITAL RADIOGRAPHY SYSTEM WITH FLAT PANEL DETECTOR
Toshiba America Medical Systems, In.C, Oct 26, 2005