ClearanceBrief
Recall Z-1516-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

GE Healthcare

GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

DateSep 29, 2008 (initiated Oct 21, 2007)
StatusTerminated
ReasonIncorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c
Root causeNonconforming Material/Component
ActionGE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed a
Quantity146 total
DistributionWorldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042053INNOVA 4100, INNOVA 3100
Ge Medical Systems, Inc., Aug 26, 2004