Recall Z-0720-2008 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Inc

Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System); Linear Accelerator with RTP Exchange v6.2, v6.6, & v8.0, All Models; Varian Medical Systems, Palo Alto, CA

DateFeb 15, 2008 (initiated Oct 2, 2007)
StatusTerminated
ReasonIncorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the gantry will not rotate, resulting in delivery to a single location rather than through the gantry arc.
Root causeSoftware design
ActionAn Urgent Medical Device Correction letter was distributed October 2, 2007, to all direct consignees, informing them that the ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery will not be recognized properly by the System. The firm provided corrective instructions and informed consignees that a product modification will be developed to resolve the issue. A Varian Service Representative will be scheduling a visit to upgrade the application.
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001643VARIS 1.4G
Varian Medical Systems, Inc., Jun 23, 2000