Recall Z-0151-2008 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

Intuitive brand da Vinci S Surgical System, Endoscopic Instrument Control System, Model IS1200 with Isolation Transformer (P/N: 952012) and Vision Cart (Model VS1000), Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA

DateFeb 22, 2008 (initiated Oct 4, 2007)
StatusTerminated
ReasonUnder-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that m
Root causeMixed-up of materials/components
ActionA product recall was initiated on October 4, 2007. Direct and Distributor consignees were sent an Important Product Notice, via electronic mail on October 5, 2007. The firm plans on sending a Field Service Engineer to inspect the system and make any necessary corrections.
Quantity63 total units
DistributionWorldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, LA, MA, MD, MN, NC, NJ, NV, OH, PA, PR, SD, TN, TX, WA, and WI, and countries of Belgium, Ireland, Italy, Korea, Russia, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021036INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS1000
Intuitive Surgical, Inc., Jun 26, 2002
K050369INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS2000
Intuitive Surgical, Inc., Apr 29, 2005
K063220DA VINCI S SURGICAL SYSTEM-V1.1, MODEL IS2000
Intuitive Surgical, Inc., Dec 1, 2006
K990144INTUITIVE SURGICAL ENDOSCOPIC INSTRUMENTS, INTUITIVE SURGICAL ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Intuitive Surgical, Inc., Jul 11, 2000