Recall Z-0266-2008 ยท HTD Forceps
Telefelx Medical
Pilling Center-Action Forceps with Insulated Shafts- Serrated Grasping Jaws, Catalog Number: 505015, Non Sterile, Rx Only, (ALLIG GRASPING FCPS 30 CM INSUL), Manufactured for : Teleflex Medical, Research Triangle Park, NC 27709 USA
| Date | Nov 29, 2007 (initiated Oct 31, 2007) |
| Status | Terminated |
| Reason | The specified insulation is not present on the shaft of the devices, leaving a risk of arcing and/or damage to surrounding tissue. |
| Root cause | Process control |
| Action | Consignees were notified by an Urgent Medical Device Recall letter on/about 11/1/2007. They were instructed to immediately cease use and distribution of the affected product and to return them to Teleflex Medical, Durham, NC. In addition, each were instructed to forward the letter if they had further distribution into institutional settings or to homecare patients and retrieve relevant product form those locations. An Acknowledgement & Stock Status Form was enclosed to be completed and faxed back to Teleflex Medical. |
| Quantity | 73 |
| Distribution | Worldwide: USA including states of AL, AR, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, NV, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV, and countries of Canada, France New Zealand and Singapore. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.