ClearanceBrief
Recall Z-0298-2008 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

Siemens, SONOLINE Antares Ultrasound System, 5.0 Systems with software version 200.0.054, P/N 05936518 and 10037593 , Siemens, Ultrasound Division, Mountain View, CA 94039

DateJan 9, 2008 (initiated Jul 10, 2007)
StatusTerminated
ReasonImage error: A software problem results in on-screen indications (graphics) that lead the user to believe that the patient's right and left and the transducer orientation are oriented the same. This is not the case and the actual patient's right and left are reversed from the on-screen indicators.
Root causeSoftware design
ActionThe firm sent a recall letter, Urgent Medical Device Notification, to consignees on 07/10/2007, to be followed by a service visit by Siemens personnel for a mandatory software update to be installed. Short term the users were instructed to avoid increasing the SieClear compounding levels to an Advanced SieClear compounding setting after an L/R invert action with Biopsy guidelines on-screen.
Quantity418 units (Z-0297-0298-2008)
DistributionWorldwide: Product was distributed to medical facilities Nationwide including the states of OH, FL, IL, NY, CA, NH, TX, NJ, PA, MD, MO, GA, WA, WI, NE, WV, HI, MA, SC, MI, VA, AZ, UT, NH, RI, KY, SD, and IN, and internationally to Canada, Germany, Australia, Sweden, Saudi Arabia, Chile, South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063803ACUSON ANTARES DIAGNOSTIC ULTRASOUND SYSTEM
Siemens Medical Solutions USA, Inc., Ultrasound DI, Jan 5, 2007