Recall Z-0169-2008 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
PRIMUS Plus,Medical charge-particle radiation therapy system (Medical Digital Linear Accelerator),MODEL NUMBER: 4504200,MANUFACTURED BY: Siemens Medical Solutions USA, Inc, 4040 Nelson Ave, Concord, CA 94520
| Date | Nov 9, 2007 (initiated Mar 22, 2007) |
| Status | Terminated |
| Reason | Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment plans. A second reason is an unplanned Dosimetry calibration data set can be erroneously copied during "Service Mode." |
| Root cause | Software design |
| Action | Distribution of Update Instructions TH007/07/S was sent to all affected sites on March 26, 2007. This was intended to inform users that Service Mode is only for use by authorized personnel, and that unauthorized and/or incorrect softpot parameter changes performed in Service Mode can result in the risk of patient mistreatment. Current control system software versions do not allow unplanned Dosimetry calibration data sets to be erroneously copied during "Service Mode." |
| Quantity | 687 units |
| Distribution | Worldwide:USA including states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K993425 | PRIMUS Siemens Medical Solutions USA, Inc., Nov 26, 1999 |