Recall Z-0287-2008 ยท HQF Laser, Ophthalmic

Ellex USA

Ellex Integre Duo, Model No: LP1RG - Ellex Medical PTY, LTD 82 Gilbert Street, Adelaide, S.A. 5000, AUSTRALIA. (The Integre Duo is a dual-wavelength surgical photocagulator laser intended for use in a clinic, outpatient clinic, or a retinal specialist's office. )

DateDec 20, 2007 (initiated Jun 13, 2007)
StatusTerminated
ReasonElectric Shock from Power Supply: The power supply has a defect which can cause an electrical shock to user.
Root causeDevice Design
ActionEllex, Inc. called each customer on 6/13/07, about the problem (defect which can cause electric shock) and advised that they would be performing modification of the power supply in the laser system and scheduling a field service visit to complete the work. Upon completion of the modification, the service technician completed the service record and left a copy with each client and informed them of the completed work. A copy of each completed service record was sent to Ellex Medical Pty. Ltd. On 9/27/07, Ellex sent a letter to all consignees, informing them that all modifications by service t
Quantity1
DistributionWorldwide Distribution - USA including states of WI, PA, CO, CA, NC, NY, SC, MD, and MS, and countries of Canada and Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052777LASEREX DUO FAMILY OF DUAL WAVELENGTH OPHTHALMIC LASERS INCLUDING MODELS INTEGRE DUO LP1RG-S, LP1RG-D
Ellex Medical Pty, Ltd., Mar 16, 2006