ClearanceBrief
Recall Z-0540-2008 ยท KPA Slide Stainer, Automated

Ventana Medical Systems Inc

Discovery, Slide Staining System, Catalog Number: N750-DIS-FS, Ventana Medical Systems Inc, Tucson, AZ

DateJan 24, 2008 (initiated Sep 29, 2006)
StatusTerminated
ReasonLeaking Carboys/Safety Updates:The Ventana Staining Platform includes one or more carboys that may have the potential to leak if the spigot is either improperly installed or not fully closed.
Root causeDevice Design
ActionThe firm has mailed three customer letters, a) the first letter dated 9/30/2005 describes the upgraded Voila CD that contains safety information; b) a second letter, dated 6/27/2006, identified as a "Customer Alert" was mailed describing an upgrade to the software system version 9.3/10.3 and the CD that includes safety improvements, c) the third letter dated 9/29/2006 and identified as a Safety Alert advises that as part of their effort to continuously improve laboratory safety, they are updating the labeling and instructions for use regarding installation and cleaning for the carboy as the Ve
QuantityUSA - 1,712 units; Foreign - 1,159 units (total for recalls-Z-0537-0541-2008)
DistributionWorldwide; USA, Australia, China, Canada, Germany, Great Britain, France, Italy, Hungary, Singapore, Korea, Taiwan, Spain, Poland, Greece, Austria, Sweden, Japan, Israel, Denmark, Turkey, Finland, Saudi Arabia, The Netherlands, Norway, Slovenia, Ireland, and Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.