ClearanceBrief
Recall Z-0281-2008 ยท KPQ System, Simulation, Radiation Therapy

Siemens Medical Solutions USA, Inc

Si emans Simiae 3000 Cassette Holder Electronic Imaging device (Image Identifier), with CT Part No: 19 57-161 or without CT Part No. 19 57 153; Siemens Medical Solutions USA, Concord, CA 94520

DateJan 31, 2008 (initiated Feb 27, 2007)
StatusTerminated
ReasonLoose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Root causeLabeling design
ActionCustomers were sent a Customer Safety Advisory Notice on February 27, 2007, notifying them that the mounting screws on the cassette holders on the SIMVIEW 3000 and NT Simulators, as well as some MEVASIM Simulators, could loosen and can cause damage to the system, or personal injury. The firm requested that they check of tightness on the mounting hardware. A STOP Modification notice was issued on 06/01/2007, to all affected regional service organizations that are responsible for implementing updates due to new information. Updated letters and User instructions will be released in August 2007
Quantity287 units total, all recalls Z-0281-285-2008
DistributionWorldwide Distribution - USA including states of IN, VA, MA, VT, NJ, NY, PA, DC, NC, SC, GA, AL, TN, MI, KY, OH, IN MI, MN, ND, MT, IL, MO, KS, NE, AR, OK, TX, AZ, CA, SD, LA, Puerto Rico, and countries of Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Colombi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K946053SIMVIEW(TM) 3000
Siemens Medical Solutions USA, Inc., Mar 31, 1995