Recall Z-0303-2008 ยท GCJ Laparoscope, General & Plastic Surgery

Stryker Communications

Extron Electronics OEM, CAB, DVI-I to DVI (3ft.); Part #: 43-112-01 (Stryker Part #: 0100-224-108); Distributed by Stryker Communications, Flower Mound, TX 75028. (These cables are commonly used with the Switchpoint Infinity II)

DateJan 12, 2008 (initiated Oct 3, 2007)
StatusTerminated
ReasonIntermitten or Complete Loss of Signal: out-of-specification DVI cables causing intermitten or complete loss of signal on monitors used in surgery.
Root causeNonconforming Material/Component
ActionFirm notified consignees of recall via letter (Urgent Device Recall) on 09/21/07. Consignees were notified that a customer service representative would visit each facility to replace out-of-spec cables with properly shielded cables at no charge.
Quantity1,653 cables.
DistributionWorldwide: USA and Australia, Belgium, Canada, France, Germany, Great Britain, Hong Kong, Japan, Korea, Mexico, Netherlands, New Zealand, Portugal, South Africa, Spain, Sweden, Switzerland, Taiwan, and The U.A.E.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033132STRYKER SWITCHPOINT INFINITY
Stryker Communications Corp., Sep 2, 2004