Recall Z-0294-2008 ยท JZL

Medtronic Navigation, Inc

Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)

DateNov 27, 2007 (initiated Oct 17, 2007)
StatusTerminated
ReasonReversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
Root causeSoftware change control
ActionMedtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
Quantity7 units
DistributionWorldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050996O-ARM IMAGING SYSTEM
Breakaway Imaging, LLC, May 5, 2005