Recall Z-0170-2008 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare

Signa MFO/I Magnetic Resonance System, Model Numbers : 2210339, 2262700, 2276937, 2282853, 2282860, 2282880, 2352129, 2388781, 5105123, 5113339, 2138300-32, 2189500-3, 2198999-4, 2259620-4, 2262700-2, 2269512-2, 2286849-2, 2352129-2, 2377062-5, 2377062-8, Manufactured by GE Hangwei Medical Systems Co., LTD, Beijing PR, China

DateNov 7, 2007 (initiated Dec 31, 2006)
StatusTerminated
ReasonIncorrect patient positioning; when moving the patient into the bore may lead to wrist and hand injury.
Root causeLabeling design
ActionField Modification Instruction : FMI 67631 (DEC 2006) was communicated directly to the GE Healthcare service engineers who provide the modification at the customer site. Corrections were performed in DEC 2006
Quantity163
DistributionWorldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, FL, ID, IL, IN, KS, KY, LA, MD, MO, MS, MT, NJ, NY, OH, OK, PA, SD, VA, and WA, and countries of United Arab Emirats, Argentina, Belgium, Brazil China, Ivery Cost, Germany, Spain, France, United Kingdom, Hungary, Italy, Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002611SIGNA MFO/I MAGNETIC RESONANCE SYSTEM
GE Medical Systems, Oct 31, 2000