Recall Z-0477-2008 ยท DTN Reservoir, Blood, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic Blood Collection Reservoir, 40 micron,Catalog/Model # EL240, Medtronic, Inc., Minneapolis, MN 55432-5804

DateJan 26, 2008 (initiated Oct 1, 2007)
StatusTerminated
ReasonUnnaproved Component Resin: Affinity NT Integrated CVR Membrane Oxygenators and EL 240 Autotransfusion Blood Collection Reservoirs contain components made from an unapproved resin
Root causeNonconforming Material/Component
ActionMedtronic sales force was notified and initiated calls with cutomers on 10/1/07. Additionally, certified Urgent Medical Device Recall letters dated 10/5/2007, were sent to the 32 customers. The letter explained the situaiton and requested that the impacted product be returned for replacement or credit. Medtronic is assessing patient use risks and will report those when determined. Standard post-surgical monitoring is advised at the present by Medtronic .
Quantity413
DistributionNationwide including the states of IL, CT, OK, OH, MN, FL, NY, KS, MO, LA, PA, IN, CA, MI, KY, SC, NE, and AR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930154EL2-SERIES BLOOD COLLECTION REERVOIR
Electromedics, Inc., Jul 1, 1993