Recall Z-0361-2008 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Plus Orthopedics USA

VKS PE Tibia Insert Ultra-Congruent; Left, Size 2, 12mm; Model Number: 0214612; Plus Orthopedics, San Diego, CA 92121

DateDec 8, 2007 (initiated Nov 13, 2006)
StatusTerminated
ReasonLabeling Error: The VKS PE Ultra-congruent insert package, may actually contain a Standard PE insert.
Root causeEmployee error
ActionConsignees were notified by telephone on 11/13/2006 to check their inventory and return any affected units. On 11/28/06 Physician Notifications and Patient Assessment forms were sent to the two physicians who had implanted a total of 4 devices.
Quantity10
DistributionNationwide: USA including states of ID, NY, VA & WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022204VKS KNEE SYSTEM
Plus Orthopedics, Oct 18, 2002