Recall Z-0359-2008 ยท DXT Injector And Syringe, Angiographic

Acist Medical Systems

ACIST Multi-Use Syringe Kit, Model A2000; SKU # : 014612; labeled as STERILE; Manufactured: Acist Medical Systems, Inc., 7450 Flying Cloud Drive, Suite 150, Eden Prairie, MN 55344, USA. Catalogue No: 800061-003. (The multiple use A20000 syringe kit is comprised of a terminally sterilized syringe barrel and a contrast spike assembly)

DateDec 8, 2007 (initiated Oct 3, 2007)
StatusTerminated
ReasonSyringe Fracture: the syringe may be fractured, causing contrast to be sprayed from syringe fracture.
Root causeProcess design
ActionThe Urgent Medical Device Recall letter was sent by certified mail to all U.S. base customers on 10/3,4/07. Customers were advised to stop distribution of the product and immediately cease use of the product. Recalled product is to be returned using the FedEx Airbill that was enclosed with the recall letter. Customers were to complete the enclosed "Device Recall Response" form even if they did not have any impacted kits in inventory.
Quantity177,050 Kits
DistributionNationwide Distribution - including states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984231ANGIOGRAPHIC CONTRAST MANAGEMENT SYSTEM, MODEL PART A AND PART B
Invasatec, May 31, 2000