Recall Z-0446-2008 ยท CBH Device, Fixation, Tracheal Tube

Dale Medical Products, Inc.

Dale 240 Blue (Adults) Tracheostomy Tube Holder, Product Number: H84102401, Dale Medical Products, Inc., Plainville, MA 02762

DateJan 11, 2008 (initiated Sep 27, 2007)
StatusTerminated
ReasonRevised Labeling : Instructions For use: "Do not trim fastener hook tabs. " Trimming can cause tabs to disengage at the trach plate
Root causeLabeling design
ActionDale Medical notified Distributors by letter dated 9/27/07, identified as an Urgent Medical Device Correction Letter. The letter requested distributors to notify all their customers by letter. The Correction Letter reminded users to take notice of new labeling " Do not trim fasterner hook tabs"
Quantity600,000 units
DistributionWorldwide-USA and countries of Canada, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Italy, The Netherlands, Norway, The UK, Spain, Switzerland, Australia, Hong Kong,India, Japan, Korea, Malaysia, New Zealand, The Philippines,Singapore, Taiwan, ROC, Brazil, Bermuda, Chile, Co
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.