Recall Z-0280-2008 ยท DSI Detector And Alarm, Arrhythmia

General Electric Medical Systems Information Technology

GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. (CIC Pro v4.1). GE Medical Systems Information Technologies, Inc., Milwaukee, Wisconsin (System hardware consists of a processing unit and a monitor)

DateFeb 1, 2008 (initiated Jan 24, 2007)
StatusTerminated
ReasonTwo separate issues may occur with operating The CiC Pro v4.1 System with Telemetry. 1) Experience a loss of audible alarms when operating system memory resource reaches approximately 170MG. This will occur if CIC Pro continuously runs for average of 160 day, and 2) System may continuously reboot upon start up in an attempt to correct the issue.
Root causeSoftware design
ActionGE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07. As a Short Term Solution, GE recommends that user reboot of system according to instructions and precautions in Service Manual at least once every three months. Also, take the necessary precautions to ensure that system is shut down properly. This will decrease the likelihood of a continuous reboot. If system continues to reboot, GE suggest that user contact Technical Support at 1-800-588-7044 (Press Option 2) and ask for Technical Support. As for a Long Term Solution, GE will provide
Quantity6,217 Units (1,549 OUS, 4,668 within US)
DistributionWorldwide Distribution - USA (excluding DE, HI, ID, MT, ND, SD, VT), Peurto Rico, Washington DC and including the countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungaria, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Mal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032370CIC PRO CLINICAL INFORMATION CENTER CENTRAL STATION
Ge Medical Systems Information Technologies, Aug 13, 2003