Recall Z-0301-2008 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Abbott Diagnostics CELL-DYN Sapphire Hemoglobin Reagent Analyzer Syringes, List Number 08H49-02, Abbott Laboratories, Abbott Park, IL (the syringe is packaged in a white box and sealed with a label wrapped around the center of the box-The main analyzer list number is 08H00-01, )

DateJan 4, 2008 (initiated Jul 25, 2007)
StatusTerminated
ReasonSyringe Failure: Insufficient lubrication of the plunger tip may cause syringes to fail at installation or shortly thereafter
Root causeNonconforming Material/Component
ActionThis Recall was initiated 07/24/07 via a Product Recall letter-Immediate Action Required-instructing users to discontinue use and destroy any units under recall. Replacements may be obtained by contacting the local Customer Support Center at 1-877-4Abbott
Quantity302 units
DistributionWorldwide Distribution: USA including states of AR, AZ, CA, CT, FL, GA, ID, KY, LA, MD, MI, NC, ND, NV, NY, OH, OK, PA, SC, TX, UT, VA, and W I (4 distributors and 50 consignees)and countries of Australia, Brazil, Chile, Germany, and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051215CELL-DYN SAPPHIRE SYSTEM
Abbott Laboratories, Jul 21, 2005