Recall Z-0257-2008 ยท JQP Calculator/Data Processing Module, For Clinical Use

Abbott Laboratories, Inc

Abbott ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; Abbott Laboratories, Abbott Park, IL 60064

DateNov 24, 2007 (initiated Oct 10, 2007)
StatusTerminated
ReasonSample problems; Five issues identified related to software utilized by the ACCELERATOR APS System Input/Output Module, to include: Under three specific conditions, the post-aspiration radio-frequency identification (RF ID) read of the tube carrier is not performed and the APS work cell may not appropriately generate Sample Presentation/Sample Queue errors with APS software version 1.1.1 or earlie
Root causeSoftware design
ActionFirm notified consignee via letter on 11/10/07. Consignee notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Quantity1 unit.
DistributionNationwide: including state of UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.