ClearanceBrief
Recall Z-0663-2008 ยท MQF Catheter, Assisted Reproduction

Fertility Technology Resources, Inc.

Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 20 cm, Sterile, Single Use Only, Order #: ftr 220, Manufactured by Fertility Technology Resources, Inc., Marietta, GA 30062

DateJan 19, 2008 (initiated Sep 13, 2007)
StatusTerminated
ReasonExpiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance.
Root causeProcess design
ActionAll customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.
Quantitynot reported
DistributionWorldwide: USA including states of CA, CO, FL, GA, LA, MD, ME, MI, MS, NC, NH, NJ, NY, OH, TN, and WA and The Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041470MORTON IUI CATHETERS
Fertility Technology Resources, Inc., Nov 22, 2004