Recall Z-0662-2008 ยท MQF Catheter, Assisted Reproduction

Fertility Technology Resources, Inc.

Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 23 CM, Sterile, Single Use Only, Order #: ft r 223, Manufactured by Fertility Technology Resources, Inc., Marietta, GA 30062

DateJan 19, 2008 (initiated Sep 13, 2007)
StatusTerminated
ReasonExpiration Date for Sterility; The 5 year dating could not be validated/supported for sterility assurance.
Root causeProcess design
ActionAll customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.
Quantitynot reported
DistributionWorldwide: USA including states of CA, CO, FL, GA, LA, MD, ME, MI, MS, NC, NH, NJ, NY, OH, TN, and WA and The Netherlands.