Recall Z-0934-2008 ยท JPA System, Multipurpose For In Vitro Coagulation Studies

Inotech Biosystems Intl., Inc

Inotech brand Heparin Adsorbant, Catalog #: HA-1, in 10mm x 75mm polystyrene tubes, each containing 70mg of Heparin Adsorbent, 10 tubes per foil lined pouch, 10 foil pouches/box, Inotech Biosystems International, Inc., Rockville, MD 20855

DateFeb 7, 2008 (initiated Aug 20, 2007)
StatusTerminated
ReasonErroneous Coagulation Tests
Root causeLabeling design
ActionThe recalling firm notified consignees by phone on 08/20/07 and provided instructions for increasing the speed of centrifugation from 1,500 x g to 2,500 x g and decanting the supernatant aliquot carefully so as to not draw in resin fibers that may be stuck on the test tube walls during sample preparation for testing in the coagulation studies. The firm updated their recall notification on 10/01/07 as an Urgent Recall/Correction to alert healthcare practitioners to not use "lupus sensitive reagents" with the heparin absorbant product and that Platelet Factor 3 and any other negatively charged m
Quantity37,370 tubes
DistributionWorldwide; USA to Hospitals and Clinical Laboratories and to one domestic distributor for export to Canada.