Recall Z-0148-2008 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Ellex Inc.

Ellex, Eye Cubed, I 3 System-ABD Ultrasound Version 3.0 with Software Version 3.0 and Biometry A-scan modality, Ellex Innovative Imaging, Sacramento, CA 95827

DateDec 28, 2007 (initiated Sep 25, 2007)
StatusTerminated
ReasonIncorrect IOL calculations: software behavior with Version 3.0 Eye Cubed Ultrasound systems running Biometry A-scan enabled Version 3.0.0 software can cause incorrect IOL calculations.
Root causeSoftware design
ActionA Global Customer Service Bulletin (GCSB)-Urgent Notification of Software Correction was issued on 25 September 2007, which provided short term workaround directions pending a free software undate promised in the near future.
Quantity59
DistributionWorldwide; including Canada and Mexico. U.S.A. distribution included states of AL, AZ, CA, CT, FL, GA, IL, MA, ME, MI, MN, NC, NE, NJ, NM, NY, OH, OR, PA, SC, TN, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041947I3 SYSTEM-ABD DIAGNOSTIC ULTRASOUND
Innovative Imaging, Inc., Aug 16, 2004