Recall Z-0288-2008 ยท JIX Calibrator, Multi-Analyte Mixture

Abbott Laboratories Inc.

Abbott Clinical Chemistry Aeroset/Architect (c Systems)Bilirubin Calibrator, List Number (LN) 1E66-03 & List Number (LN) 1E66-04, Abbott Laboratories, Abbott Park, IL 60064

DateAug 10, 2007
StatusTerminated
ReasonIncorrect Calibrator values: changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the standard.
Root causeMaterial/Component Contamination
ActionAn August 10, 2007, Product Correction letter was provided to all ARCHITECT cSystem and AEROSET System customers. The letter advises that changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the NIST standard 916a. Revised Bilirubin values for the calibrator lot numbers in this recall are attached to the recall notice. Six necessary actions were listed in the recall notice as: 1) Identify the lot numbers of Clinical Chemistry Bilirubin Calibrator(s) currently in use. 2) Discard the value sheet and replace with the
Quantity7,033 units total worldwide
DistributionWorldwide: USA, Canada, Mexico, Guatemala, Puerto Rico, Germany, Hong Kong, Singapore, Australia, New Zealand, Jamaica, Venezuela, Columbia, Ecuador, Chile, Uruguay, Argentina, Brazil, Thailand, Trinidad & Tobago, Bahamas, Panama, Cayman Islands, and El Salvador
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981706AMM CAL, BIL CAL, CO2 CAL, HDL CAL, IRON/MG CAL, LDL CAL, MC CAL, UPRO CAL, AND ISE CALIBRATOR LEVELS 1 AND 2
Abbott Laboratories, Jul 9, 1998