Recall Z-0126-2008 ยท GWQ Full-Montage Standard Electroencephalograph

Nicolet Biomedical Div of Viasys Healthcare

Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711

DateJan 3, 2008 (initiated Sep 11, 2007)
StatusTerminated
ReasonDisplay lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.
Root causeComponent design/selection
ActionCertified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.
Quantity33
DistributionWorldwide Dsitribution - USA, Germany, Netherlands, Kuwait, Japan, Ireland, and China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021185NERVUS MONITOR
Taugagreining HF, Jul 12, 2002