ClearanceBrief
Recall Z-0270-2008 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA Inc

Siemens ECAT ART, Catalog No.. 3545584, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc., 810 Innovation Drive, Knoxville, TN 37932 USA, TEL: 865-218-2534

DateNov 29, 2007 (initiated Jul 23, 2007)
StatusTerminated
ReasonIncorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
Root causeSoftware design
ActionSiemans sent an Urgent Device Correction letter, dated June 15, 2007, to all affected consignees informing the user of the problem and requesting that they discontinue the practice of running all transmissions followed by all emissions until a corrective software patch is installed. The corrective software patch is expected to be installed by October 2007.
Quantity43 units
DistributionWorldwide Distribution - USA and Countries of Argentina, Austria, Belgium, Brazil, Canada, Switzerland, Chile, China, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Italy, Japan, South Korea, Netherlands, Norway, Saudi Arabia, Sweden, Slovak Republic, Turkey, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002584ECAT ACCEL PET SCANNER
Cti Pet Systems, Inc., Nov 16, 2000
K003541DENLASER, MODEL 800
CAO Group, Inc., Jan 25, 2001
K940478ECAT PRT
Cti Pet Systems, Inc., Mar 22, 1994
K962797ECAT EXACT & ECAT EXACT HR+
Cti Pet Systems, Inc., Oct 15, 1996