ClearanceBrief
Recall Z-0348-2008 ยท MBI Fastener, Fixation, Nondegradable, Soft Tissue

Axya Medical

Arthrocare 3mm ParaFix Anchor, single strand of Magnum Wire,Part Number: 22-5010, Axya Medical, Inc., Beverly, MA 01915

DateDec 8, 2007 (initiated Sep 27, 2007)
StatusTerminated
ReasonSterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root causePackage design/selection
ActionAxya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity338 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051983MODEL 3000 AXYALOOP BIO-ABSORBABLE BONE ANCHOR
Axya Medical, Inc., Aug 5, 2005
K052491AXYA MODEL 6500 AXYALOOP SELF-TAPPING BONE ANCHOR
Axya Medical, Inc., Oct 7, 2005
K060970AXYALOOP NONABSORBABLE BONE ANCHOR
Axya Medical, Inc., Jul 5, 2006