ClearanceBrief
Recall Z-0330-2008 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

Allura Xper FD20/10, Release 1.2, Philips Medical Systems, Bothell, WA 98021

DateDec 7, 2007 (initiated Sep 24, 2007)
StatusTerminated
ReasonInaccurate anatomical measurements when using the optional SmartMask feature
Root causeDevice Design
ActionThe recalling firm sent customers an Urgent Device Correction notification letter dated 09/21/07 via certified mail - return receipt requested. This notification provided a short term work around for the device defect. For a permanent fix, the recalling firm is developing software update to check for mismatch at the start of a subtracted fluoroscopy run, and will contact customers to schedule an install. Installation is expected to begin 10/2007 and be completed 06/2008.
Quantity53 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003