ClearanceBrief
Recall Z-0429-2008 ยท NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing

Depuy Orthopaedics, Inc.

Depuy Preservation Uni Knee All Poly Tibia, RM/LL, Sz 4, 11.5 mm, sterile; Product Code/REF 1498-11-410.

DateDec 6, 2007 (initiated Sep 27, 2007)
StatusTerminated
ReasonAn incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root causePackaging process control
ActionConsignee hospitals and implanting physicians were notified of the recall via an Urgent Information-Recall Notice letter and Dear Doctor letters dated 9/27/07. The hospitals were instructed to determine whether they had any of the recalled products and to contact their sales representative to pick up the product. Depuy sales representatives were notified to collect the products from consignee hospitals via email on 10/1/07. The decision to discuss this product defect with patients who had already been implanted was left to the implanting physicians.
Quantity9
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010810DEPUY PRESERVATION UNICONDYLAR KNEE PROSTHESIS
DePuy Orthopaedics, Inc., Apr 18, 2001