Recall Z-0245-2008 ยท JHI Visual, Pregnancy Hcg, Prescription Use
Applied Biotech Inc
Clearview hCG Combo Waived for urine & Moderate for serum, Wampole Laboratories, ABI List No/Product No. 6026KCV; Contents 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA
| Date | Nov 21, 2007 (initiated Sep 28, 2006) |
| Status | Terminated |
| Reason | Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product. |
| Root cause | Process design |
| Action | A Product Recall Notification dated 9/28/2006, was provided via fax, email or direct mail to all primary consignees. Customers were instructed to block all stock of affected lots in their warehouses, contact their primary consignees and advise them to discontinue use immediately and dispose of any unused materials; using a fax back form to communicate to the recalling firm the quantity blocked and disposed of so that arrangements could be made for providing replacement product |
| Quantity | 3,238 Kits (97,140 tests) |
| Distribution | Worldwide: USA and Canada |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.