Recall Z-0295-2008 ยท JBS Timer, General Laboratory

Nova Biomedical Corporation

Nova 8 Analyzer Calibrator Pack (Catalog/REF # 15196), Nova Biomedical, Waltham, MA

DateJan 17, 2008 (initiated Sep 17, 2007)
StatusTerminated
ReasonElevated Normalized Ionized Calcium (nCa) and Normalized Ionized Magnesium (nMg) calculated values on patient samples
Root causeDevice Design
ActionNova issued a Customer Advisory Notice notification by phone on 9/17/07 to contact customers using a script . If the facility reports Normalized Ionized Calcium or Normalized Ionized Magnesium, Nova Biomedical will replace existing stock of Nova 8 calibrator packs. If the facility does not report Normalized Ionized Calcium or Normalized Ionized Magnesium, the account can continue to use the existing lot of calibrator packs.This contact will be documented within the Nova Biomedical complaint database. Intemational Customer Advisory Notice being sent to Nova Biomedical Subsidiariies and Distri
Quantity2,999 units
DistributionWorldwide, USA, Canada, Argentina, Bangladesh, Bolivia, China, Czech Republic, Denmark, Ecuador, Egypt, Finland, France,Germany,Greece, Hungary, India, Indonesia, Ireland,Italy, Japan, Korea, Mexico, Peru, Poland, Scotland, South Africa, Spain, Sri Lanka. Switzerland, Syria ,Taiwan, Turkey and The U
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.