Recall Z-0120-2008 ยท JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

AmniSure International LLC

AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138

DateJan 19, 2008 (initiated Oct 3, 2007)
StatusTerminated
ReasonWeak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"
Root causeDevice Design
ActionAmniSure notified users by Medical Device Correction (Recall) letter dated October 1, 2007 issued via US Mail. The letter provides additional labeling instructions and a checklist for trouble shooting, including a diluation procedure that allows users to determine and solve the problem of the hook effect with this device.
Quantity319,976 units
DistributionWorldwide, USA, Australia, Austria, Italy, Korea, Bulgaria, Russia, Spain, Kuwait, Switzerland, Luexemburg, Ireland, Denmark, Netherlands, Jordan, Israel, Chile, UK, Singapore, Basil, New Zealand, Turkey, France, Greece, and Equador