ClearanceBrief
Recall Z-0313-2008 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc

COULTER LH780 Analyzer , Part Number: 723585, Beckman Coulter, Brea, CA. (Hematology analyzer)

DateDec 22, 2007 (initiated Apr 18, 2007)
StatusTerminated
ReasonPatient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th
Root causeSoftware design
ActionProduct Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system
Quantity27 units in the US; 14 units in Canada
DistributionWorld wide: USA and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061616COULTER LH 780 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Oct 4, 2006