Recall Z-0304-2008 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

c.cam Gamma Camera, Model No: 9VIR1200; Tomography System (Emission Computed); Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203; Made in Denmark

DateDec 19, 2007 (initiated Sep 17, 2007)
StatusTerminated
ReasonUnintended Motion: A hardware fault in the c.cam system can cause the system to experience unintended motion, thereby causing a collision hazard.
Root causeSoftware design
ActionUrgent Field Correction Recall letters dated 9/11/07, on Siemens letter head were sent to the end users via U.S. certified mail, informing them of the potential risk of unintended motion of the detector head resulting in a collision hazard. The end users were warned not to press any of the detection motion buttons while simultaneously holding the collision over ride button on the c.cam hand controller, and were provided steps to take if the system stops during an acquisition and displays the Collision Error message. They were told that a Siemens service representative will contact them withi
Quantity486 units
DistributionWorldwide: USA including Puerto Rico and Brazil, Canada, France, Germany, Great Britain, Israel, The Netherlands, and Portugal.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031825C.CAM
Danish Diagnostic Development A/S, Jun 20, 2003