Recall Z-0188-2008 ยท MNI Orthosis, Spinal Pedicle Fixation

Allez Spine, LLC

Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 45, Catalog Number LG-PR0845, Allez Spine, LLC., Irvine, CA

DateMar 11, 2008 (initiated Jul 30, 2007)
StatusTerminated
ReasonScrew Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
Root causeDevice Design
ActionOn July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users wa
Quantity325
DistributionNationwide: including the states of AZ, CA, CO, GA, NV, TX and UT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050060LAGUNA PEDICLE SCREW SYSTEM
Allez Spine, LLC, May 4, 2005