Recall Z-0244-2008 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Applied Biotech Inc

Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA

DateNov 21, 2007 (initiated Sep 28, 2006)
StatusTerminated
ReasonSensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
Root causeProcess design
ActionA Product Recall Notification dated 9/28/2006, was provided via fax, email or direct mail to all primary consignees. Customers were instructed to block all stock of affected lots in their warehouses, contact their primary consignees and advise them to discontinue use immediately and dispose of any unused materials; using a fax back form to communicate to the recalling firm the quantity blocked and disposed of so that arrangements could be made for providing replacement product
Quantity400 Kits (1,200 tests)
DistributionWorldwide: USA and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.