Recall Z-1155-2008 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

GE Healthcare Clinical Systems

GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201

DateJun 12, 2008 (initiated Sep 4, 2007)
StatusTerminated
ReasonInaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause incorrect guidelines to be displayed on the image. The incorrect guidelines are recognized by their orientation that is left/right reversed from t
Root causeSoftware change control
ActionGE sent consignees a "GE Urgent Medical Device Correction" letter dated 8/14/07. The letter informed the customers about the problem, offer a short-term workaround and informed them that a representative will schedule a visit to upgrade the software.
Quantity62
DistributionWorldwide Distribution - USA (NY & AL), Korea, China, India, Brazil, Mexico, Germany, Greece, Italy, Poland, Russia, Saudi Arabia, Spain, and Czech Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020263LOGIQ 3
General Electric Co., Feb 6, 2002