ClearanceBrief
Recall Z-0119-2008 ยท JGS Electrode, Ion Specific, Sodium

Opti Medical Systems Inc

OPTI LION E-Plus Cassettes, Model BP7507, (25 per box), In Vitro Diagnostic, OPTI Medical Systems, Inc., Roswell, Georgia 30076

DateNov 9, 2007 (initiated Jul 18, 2007)
StatusTerminated
ReasonIncorrect results/stability failure; Actual device stability studies show sodium and chloride sensors age differently than anticipated when stored at the high end of labeled storage, therefore subsequent measurements give results outside the allowed error allowance before the 6 month expiration date shown on the device.
Root causeDevice Design
ActionAn Urgent Medical Device Recall Letter dated 16 July 2007, was issued via E mail and Fax on/or about 7/18/2007 directing consignees to remove unused cassettes from inventory and destroy them. Replacement cassettes will be shipped upon receipt of information regarding destruction by calling customer service at 800-490-6784. Sub-recall by distributors was requested.
Quantity289 boxes of 25 cassettes
DistributionWorldwide Distribution-USA, Australia, Chile, China, Egypt, France, Great Britain, Guatemala, Indonesia, Italy, Korea, Lebanon, Mauritania, Morocco, Nepal, Russia, South Africa, Sweden, Tanzania, Thailand, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052027OPTI LION ELECTROLYTE ANALYZER, MODEL GD7200
Osmetech, Inc., Sep 29, 2005