Recall Z-0125-2008 ยท KWP Appliance, Fixation, Spinal Interlaminal

Warsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK

Medtronic Sofamor Danek COLORADO 2" SPINAL SYSTEM Break-Off Nut, non-sterile, RX; Catalog Number 8634111, Medtronic, Memphis, TN 38138.

DateFeb 7, 2008 (initiated Sep 10, 2007)
StatusTerminated
ReasonAssembly may be compromised: Two thread grooves are missing on the nut, which may cause damage to the bone screw and result in the need for the bone screw to be replaced.
Root causeProcess control
ActionMedtronic Sofamor Danek notified consignees by Urgent Device Recall letter dated 9/10/07 to immediately cease using the product and to return the devices with a Product Recall Questionnaire, recording quantity of devices and any patient injuries.
Quantity2,569
DistributionWorldwide Distribution - USA, Australia, Hong Kong and the Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991031COLORADO II SPINAL SYSTEM
Sofamor Danek USA,Inc., May 10, 2000