Recall Z-0124-2008 ยท DPS Electrocardiograph

Philips Medical Systems

Philips PageWriter Touch Cardiograph, Model: 860284, Philips Medical Systems, Andover, MA, 01810

DateJan 30, 2008 (initiated Sep 12, 2007)
StatusTerminated
ReasonIncorrect display: When using the 16 lead Patient Interface Module (PIM), lead tracings of V3R and V4R are reversed in the extended lead modes of Pediatric and Balanced.
Root causeSoftware change control
ActionPhilips Medical Systems notified affected customers by Customer Letter and Urgent - Device Correction Notice dated 9/12/07. The correction will be a software upgrade. In the US, the Philips Response Center will contact US customers and arrange their upgrades, either via a software download, or by sending a Field Service Engineer to the customer site to perform the upgrade. Until new software is installed, customers are being advised to use the 12 lead module, and to disable or disconnect the 16 lead connection.
Quantity266 units
DistributionWorldwide Distribution - USA, Canada, Australia, Hungary, New Zealand, Nigeria, Norway, Qatar, Taiwan, and UAE.
FDA recordOpen on fda.gov