ClearanceBrief
Recall Z-0254-2008 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Med Sys Inc

NEMIO XG Ultrasound System, Model SSA-580A , Software versions: VI .5.0, Vl .5.000A, VI .5.000B, V2.0.0,Toshiba America Medical Systems, Inc., Tustin, CA.

DateNov 17, 2007 (initiated Aug 29, 2007)
StatusTerminated
ReasonMultiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.
Root causeSoftware design
ActionUrgent Medical Device Notification letters were sent to customers on August 29, 2007. Customers were notified of the several issues related to the Nemio SSA-550A and Nemio SSA-580A recall and asked to share this information with all users of the device. Customers were informed that Toshiba is preparing software and corrected manuals for distribution to address these issues. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department.
Quantity128 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061605NEMIO XG, SAA-580A
Toshiba America Medical Systems, In.C, Jun 22, 2006