Recall Z-0101-2008 ยท MED Sterilant, Medical Devices

Steris Corporation Hopkins Facility

Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays) The malfunctioning piece of the quick connect component is adaptor #6, STERIS part number 201158. Adaptor #6 attaches to the suction port of the endoscope.

DateOct 24, 2007 (initiated Sep 12, 2007)
StatusTerminated
ReasonLoose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the attached endoscope.
Root causeComponent design/selection
ActionSTERIS has contacted affected direct accounts by phone to notify direct accounts of the recall. On 9/28/2007, STERIS will mail a recall notification to all affected direct accounts via certified mail. The notification informs direct accounts that STERIS representatives will be visiting direct accounts to retrieve affected products and provide replacements. Where direct accounts further distribute, STERIS representatives will inform the direct accounts to contact their sub-accounts and facilitate exchange of affected product for replacement product. Replacement product will be available beg
Quantity391 Quick Connects
DistributionWorldwide: United States, Australia, Bahrain, Canada, Germany, Great Britain, Hungary, Italy, Qatar, and Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.