Recall Z-0134-2008 ยท FRN Pump, Infusion

Smiths Medical MD, Inc.

Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA

DateFeb 26, 2008 (initiated Aug 14, 2007)
StatusTerminated
ReasonIncorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.
Root causeSoftware design
ActionA Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.
Quantity3 CD-ROMs
DistributionUT, MD, CO