ClearanceBrief
Recall Z-0049-2008 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Boston Scientific CRM Corp

Guidant ASSURE, Model B301, Implantable Cardioverter Defibrillator, ICD. (This model is not available in the USA) Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.

DateNov 24, 2007 (initiated Mar 16, 2007)
StatusTerminated
ReasonAtypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal extension of the ERI charge time . ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product
Root causeDevice Design
ActionA Product Update titled "ERI Charge Time Limit Extended During Mid-Life and Mid-Life Display of Replacement Indicators", dated 03/10/2007 was posted to Company's website on 03/13/2007. A CRM Product News letter was sent to physicians beginning 03/16/2007. The CRM Product News Letter gave an overview of the content of the Product Update and stated the web location for finding the complete Product Update. The Product Update identified affected devices, gave an overview of normal charge time behavior at mid-life for specific ICD and CRT-D devices. The letter also described the atypical charge
Quantity180-Recalled units only cover devices implanted prior to July 2005
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.