ClearanceBrief
Recall Z-0099-2008 ยท KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented

Encore Medical, Lp

Foundation Hip System Bipolar Modular, Size 60mm OD; Cat. #412-01-060; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.

DateNov 7, 2007 (initiated Aug 2, 2007)
StatusTerminated
ReasonDevice mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
Root causePackaging change control
ActionFirm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Quantity13 units.
DistributionWorldwide, including USA, Greece, Japan, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953510HIP JOINT FEMORAL (HEMI-HIP) METAL/POLYMER CEMENTED OR UNCEMENTED PROSTHESIS
Encore Orthopedics, Inc., Aug 9, 1995