ClearanceBrief
Recall Z-0100-2008 ยท JOH Tube Tracheostomy And Tube Cuff

Smiths Medical ASD, Inc.

Portex Perfit Percutaneous Tracheostomy Tube Replacement Kit, Sterile Ref: 536090

DateNov 7, 2007 (initiated Sep 18, 2007)
StatusTerminated
ReasonMislabeled. The product is labeled as a size 9.0 mm tracheostomy tube, but the package contains a 7.0 mm tracheostomy tube.
Root causeProcess control
ActionSmiths Medical contacted direct customers and US distributors by letter sent by UPS on Septmeber 18, 2007. US Distributors instructed to return their inventory and provide a customer list to Smiths Medical.
Quantity40 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031057PER-FIT PERCUTANEOUS DILATIONAL TRACHEOSTOMY KIT
Portex, Inc., Jul 18, 2003