Recall Z-0074-2008 ยท KWP Appliance, Fixation, Spinal Interlaminal

Medtronic Sofamor Danek USA Inc

Medtronic Sofamor Danek, TSRH SiLo 5.5 Spinal System, Sagittal Adjusting Screw Implants, Supplement Sets, REF: 8889245, Pedicle Screw Fixation System, Medtronic Sofamor Danek, Memphis, TN 38131

DateDec 13, 2007 (initiated Aug 24, 2007)
StatusTerminated
ReasonDisassembly: The movable cradle component could detach from the saddle under extreme manipulation of the associated rods prior to final tightening of the set screws.
Root causeDevice Design
ActionMedtronic sent letters, titled URGENT DEVICE RECALL, on 08/24/2007, to the Risk Managers of the hospital consignees explaining the potential for these sets to disassemble, and requesting the discontinuation of the use of and quarantine of the product. The product will be retrieved by the firm and replaced.
Quantity20 units
DistributionWorldwide Distribution - USA, The Netherlands, Australia, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052144TSRH SPINAL SYSTEM
Medtronic Sofamor Danek, Aug 31, 2005