Recall Z-0062-2008 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems

HeartStart XL Defibrillator/Monitor, Model: M4735A

DateOct 25, 2007 (initiated Sep 4, 2007)
StatusTerminated
ReasonInadequate shipping package causing damage to control boards resulting in unit failure. Failure to deliver therapy (defibrillation, pacing or cardioversion) may occur as a result of undetected damage to the PCB assembly.
Root causePackage design/selection
ActionPhilips Medical notified accounts (Distributors/Healthcare Customers) by letter dated 9/4/07. Users are requested if possible to remove the unit from service. Firm will have Field representative conduct correction.
Quantity147 boards
DistributionPuerto Rico, Canada, Argentina, Australia, Brazil, China, Domincan Republic, Czech Republic, Hong Kong, India, Japan, Korea, Macau, Singapore, Spain, South Africa, Slovenia, Sweden, Switzerland, Ireland, UK, France, Kuwait, Poland, and Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021453HEARTSTREAM XL DEFIBRILLATOR/MONITOR, MODEL #M4735A
Philips Medical Systems, Jul 3, 2002